Fetal MRI, set right in minutes.

Protocol Review Desk helps hospitals and imaging centres get consistent, high-quality fetal MRI on GE HealthCare, Philips and Siemens Healthineers scanners. It reads the protocol and the images, checks the key settings against approved references, and shows your team how to fix what’s off.

Series 01Read

Readprotocols and images.

DICOM® studies, GE HealthCare protocol folders, Philips ExamCard files, Siemens Healthineers exam archives (.exar1) and PDF printouts. Each value keeps its vendor’s meaning and says where it came from.

Series 02Compare

Compareplanned and acquired.

The protocol on the console next to what the DICOM says was acquired. Values more than 3 % and 0.05 apart are flagged, one by one.

Series 03Check

Checkagainst approved references.

Fetal heart, fetal brain, and fetal body and placenta, at 1.5 T and 3 T. Every range carries its origin, expert or literature, and a version.

Series 04Fix

Fixin each vendor’s own terms.

What’s off, the change in that vendor’s own parameter names, and what it costs in time or signal. Your team approves every change at the console.

Series 05Local

Localby default.

Files are read on your computer and nothing is uploaded. Patient data stays in the hospital. A technical tool for image quality, never a diagnostic one.

Series 06See it work · built-in example, synthetic data

From 71 to 100. In one review.

This is the desk’s own automatic review of its example case: 6 settings outside the fetal reference, each with the change in the scanner’s own words and what it costs. Approve them and watch the score.

Show the fixes for
  • Slice gap · T2 single-shot

    1 mm0mm

    Recommended 0 mm

    Geometry tab: slice gap → 0 mm (user defined); interleaved slice order

  • Averages · Cine bSSFP

    42

    Recommended 1–2

    NSA → 2

  • Averages · Cine bSSFP

    42

    Recommended 1–2

    NSA → 2

  • Field of view (read) · Cine bSSFP

    323 × 323320mm

    Recommended 250–320 mm

    Geometry tab: FOV → 320 mm

  • Reconstructed phases · Cine bSSFP

    2425

    Recommended 25–50

    Motion tab: heart phases (reconstructed) → 25

  • Reconstructed phases · Cine bSSFP

    2425

    Recommended 25–50

    Motion tab: heart phases (reconstructed) → 25

Up close · one finding, and planned against acquired

FindingFetal CMR · cine bSSFP · Philips · 1.5 T
Problem
Breath-hold per slice 7.3 s
recommended ≤ 7 s · expert
Reference
Fetal CMR v1.0 · house maximum 7 s for every breath-hold sequence
FixP
SENSE factor 1.9 → 2: breath-hold about 6.9 s (SNR about −3 % or more)
Trade-off: Lower SNR or spatial resolution.

From the validation set · worded as the desk reports it

Compare · planned vs acquiredPlannedAcquired
  • (0018,0080)Repetition Time3.6 ms3.74 msDiffers
  • (0018,0081)Echo Time1.78 ms1.87 msDiffers
  • (0018,1314)Flip Angle60 °60 °Match
  • (0018,0095)Pixel Bandwidth1624 Hz/px973 Hz/pxDiffers
  • (0018,0050)Slice Thickness5 mm5 mmMatch
  • (0018,0083)Averages4 4 Match

Synthetic example from the desk · flagged when more than 3 % and 0.05 apart

Series 07Who it’s for

For the teams behind every scan.

Radiologists, MR technologists, physicists and MR leads

Keep fetal protocols consistent across scanners, shifts and sites. When images look worse, load the case, see planned against acquired, and fix it together at the console. Check the protocol again after every upgrade or service visit.

  1. T1When images look worse
  2. T2After an upgrade or service visit
  3. T3Across scanners and sites
Approved references today
Fetal heart · fetal brain · fetal body and placenta
Next anatomies
Adult cardiac · brain · abdomen and pelvis · joints

Series 08Principles

  • ROI 01

    Your team approves everything

    The desk finds, explains and proposes. A person approves every finding and every change.

  • ROI 02

    Nothing written to the scanner

    Changes are made on the console by your team, never by the tool.

  • ROI 03

    Patient data stays on site

    Files are read on your computer. Nothing is uploaded, and there is no telemetry on content.

  • ROI 04

    Technical, not diagnostic

    It reviews how images are acquired, never what they show.

Series 09Five-minute moments

  • Q1Why is this fetal cine flagged?
  • Q2What is this parameter called on another vendor’s scanner?
  • Q3How long is this breath-hold at 140 bpm?
  • Q4What happens to time and SNR if I change this?

Series 10Bring a case

Bring a case.

A fetal study, an exam card or a protocol printout. Open the desk and see it read, checked and explained on your own computer, or try it on the built-in example first.